【Bossmen Nitrogen Control System – N2HT】
Bossmen nitrogen high-efficiency dehumidification system, equipped with various stainless steel nitrogen cabinets, conforms to international semiconductor drying and cleanliness standards. IEC-TR- 62258-3 International Semiconductor Humidity Control Specification, is the humidity standard for the storage of wafers (Wafer) and chips (Bare die) in the semiconductor industry. The specification states that wafers or chips should be stored in a nitrogen concentration of at least 99% or in a dry air environment, and the maximum humidity should not exceed 30% RH, and the minimum should not exceed 7% RH (to prevent extremely low humidity environments and no electrostatic processing will occur. cause electronic damage). In addition, the most important thing is that the storage environment cannot generate dust (particles), and the dust particles must meet the standards of ISO14644, Class 6.
Bossmen’s rapid dehumidification nitrogen control series is designed in full compliance with international standards. It not only has the reliability of dehumidification, but also the static elimination system can avoid dust generation and sticking. If combined with a high-efficiency filtration system, the problem of dust disturbance in the cabinet can also be avoided. Nitrogen energy-saving device, as long as the humidity set point is reached, the nitrogen filling will be stopped, saving nitrogen consumption by 70%.
【Bossmen Nitrogen Control System – N2HT】
Product Specifications:
N2HT nitrogen dehumidification (add oxygen concentration control function)
Humidity Control Range | 1%RH~60%RH adjustable | Static elimination | Yes, intake ion nozzle |
Humidity display range | 0.0%RH~99.9%RH | Temperature and humidity recording function | Yes, 11,900 strokes |
temperature display range | 0.0℃~ 60.0℃ | Temperature and humidity alarm function | Yes, with buzzer and flashing |
Oxygen concentration display range | 0.1%~25.0% | Filter function | According to nitrogen cleanliness, a finer filter system is optional |
Temperature Accuracy | ±1℃(0~60℃) | Offset correction function | Yes, can be set by software |
Moisture Accuracy | At 25℃±3%RH(1~60%RH) | Panel lock function | Yes, can be set by software |
Oxygen Concentration Accuracy | ±0.2% (0~25%) | Display screen size | 2.3″ LED numbers |
Inlet Pressure | Nitrogen, 3kg / cm2 ~ 7kg /cm2 | device size | W315XH337XD220mm |
power | 110V~220V / 50Hz~60Hz | Power consumption | 25W |
Can be directly upgraded with various moisture-proof cabinets or nitrogen cabinets, easy to install, fully compliant with IEC-TR-62258-3 International Humidity Control Code |